PRODUCTS

Ethylene oxide sterilization equipment

Ethylene oxide sterilizer

HTY series ethylene oxide sterilizer is a special equipment for low-temperature fumigation and sterilization of articles sealed in the sterilization room under a certain temperature, pressure, humidity and a certain action time with the mixture of pure ethylene oxide and ethylene oxide. Ethylene oxide gas is a very active chemical disinfectant, which can not only achieve broad-spectrum sterilization, but also achieve a reliable sterilization effect.

Specifications and parameters of ethylene oxide sterilizer

Model HTY-8cube
HTY-10cube HTY-20cube
Disinfection mode Low temperature
Low temperature
Low temperature
Rated power
18KW 22KW 22KW
Liner specifications length * width * height (mm) 5200*1200*1300 4200*1400*1700 5000*1800*2250
Working pressure 0--60Kpa 0--60Kpa 0--60Kpa
Adjustable working temperature 50°C ±3°C 50°C ±3°C 50°C ±3°C
Working humidity 45%‐75%
45%‐75%
45%‐75%
Sterilization time 0-99 hours adjustable 0-99 hours adjustable 0-99 hours adjustable
Material of disinfection room 304 anticorrosive stainless steel 304 anticorrosive stainless steel 304 anticorrosive stainless steel

Advantages:
1. Simple operation: one-button operation, microcomputer control, simple and convenient, digital display of sterilization process.
2. Print records: Record and print sterilization information for sterilization process, which is convenient for verification and archiving.
3. Filtration system: The air inlet is equipped with a filtration system, which can effectively filter out germs and microorganisms and avoid secondary pollution during exhaustion.

Application:
Hysteroscopy, Laparoscopy, Laryngoscope, Probe, Rigid Endoscope, Hose Endoscope, Ophthalmic Lens, Fiber Optic Cable, Cranial Pressure Sensor, Cold Therapy Probe, Prostatectomy Device, Esophageal Dilator, Electrocautery Devices, defibrillation electrodes, laser head metal, glass, silicone, Chinese medicine and western medicine, human implants, masks, DNA, etc.

Installation requirements of ethylene oxide sterilizer:
1. The ethylene oxide sterilizer must be placed in a well-ventilated place, and it must not be placed near a fire source.
2. In order to facilitate repair and regular maintenance, 51cm space should be reserved on each side (including the top) of the ethylene oxide sterilizer.
3. Special exhaust pipes should be installed and completely isolated from other exhaust pipes of the building.

Ethylene oxide safety protection principles and precautions:
1. Keep ethylene oxide sterilizers and cylinders or cans away from sources of ignition and static electricity.
2. Ethylene oxide storage should be free of ignition sources, no rotating motor, no sun, good ventilation, and the temperature is lower than 40°C, but it should not be placed in the refrigerator. Dispose of in strict accordance with the relevant national storage requirements for flammable and explosive materials.
3. Do not use too much force when administering or opening the bottle, so as not to spray the liquid.
4. Monitor the air concentration of the ethylene oxide working environment every year.
5. Ethylene oxide staff should be trained in expertise and emergency management.

Technical description of ethylene oxide sterilizer

Air replacement: Automatic ventilation, air exchange frequency and interval can be set freely
Automatic control system: One-click start to end, the whole process is automatically completed with print record
Display status: Display temperature, humidity, pressure, sterilization time, residual gas treatment time, number of times
Printer: The whole process of sterilization information record printing, in order to verify the check.
Humidity sensor: The built-in humidity sensor automatically detects the humidity in the cavity to ensure that the humidity in the cavity is not lower than 45% RH.
Pressure sensor: The inner pressure sensor automatically detects the pressure inside the cavity to ensure the accuracy and efficiency of the pressure.
Temperature sensor: The built-in temperature sensor automatically detects the cavity temperature to ensure that the cavity temperature remains within the accurate range of 50 °C±3°C.
Humidifying system: Pulse humidification in the vacuum state ensures that the humidity in the cavity is uniform and controllable, and the humidity of the sterilizing chamber should be controlled within the range of 45- 80% relative humidity.
:Residual gas treatment unit: Filtration efficiency of filter for particle diameter >=0.3 μ m,not less than 99.5%.
Category: Class II medical instruments.

Ethylene oxide sterilization equipment for sale. KMEC provide different size ethylene sterilization machine. Ethylene oxide sterilization process & equipment, Good effect, safety protection.

CONTACT US

If you're interested in our products or have any questions, please let us know. Don't hesitate to contact us!